Sigma 6000

The Sigma 6000 features primary and Secondary infusion functions, plus a stepped 12-part program feature.

Category: Tags: ,

Product Description

The Sigma 6000 features primary and Secondary infusion functions, plus a stepped 12-part program feature.
Can be programmed to require a drop sensor, or set max. and min. rates or volumes.
Keypad lock.
Memory is 3 hours after Power Off in Pri- mary Mode, permanent in Program Mode.
Use of drop sensor in the program mode will disable the Door air sensor, a useful feature for customers experiencing nuisance air alarms.
In revisions below 1.3, more than one rate must be entered into the program mode to enable pump operation.

Rate Parameters: 0.1 – 999 mL / hr.
Volume Limit Range: 0.1 – 9999 mL.
KVO: 1.0 mL / hr, or infusion rate if < 1 mL / hr. Battery Life: Fully-charged, 8 hours. (18 hour recharge) Occlusion Pressure: 12 ± 6 PSI. Memory: 3 hours in Primary Mode, permanent in Program Mode. Accuracy: ± 5%.