Baxter Colleague 3

The Baxter Colleague 3 pumps enhance the Colleague pump family with the Guardian feature to help reduce medication errors by alerting staff when programmed doses are not met within institutional limits.

Product Description

**This product has been recalled by the FDA. For additional details, please view the press release by the FDA about the recall here.**

The Baxter Colleague 3 pumps enhance the Colleague pump family with the Guardian feature to help reduce medication errors by alerting staff when programmed doses are not met within institutional limits
AC Power Requirements 100-120 VAC
60Hz Leakage Current
Less than 300 u amps earth leakage (tested per UL 2601)
External Fuses
1.6 amp Type T (time delay) 250 V
Power Cord
2.9 m (9 feet) long with integrally molded hospital grade plug
Batter Supply System
Colleague 3 Pump – 3 hours operating time at 100 mL/hr for all three channels (fully charged new batteries, 12 hours to recharge to 80% of capacity. The internal charge system recharges batteries whenever the device is connected to an AC outlet.
Range of Programmable Flow Rates
– Primary infusion on each pump channel 0.1 to 99.9 mL/hr in 0.1 mL/hr increments, 1 to 1200 mL/hr in 1 mL/hr increments
Piggyback infusion on each pump channel 0.1 to 99.9 mL/hr in 0.1 mL/hr increments, 1 to 500 mL/hr in 1 mL/hr increments
– Rate limits can be configured for values less than those stated above
Volume to be Infused
– 0.1 – 99.9 mL in 0.1 increments
– 1 – 9999 mL/hr in 1 mL increments