The ValleyLab Force 1B features monopolar and bipolar outputs, audible and visual arms, including REM(TM) Fault Indicator.
ValleyLab Force 1B
Product Description
The ValleyLab Force 1B features monopolar and bipolar outputs, audible and visual arms, including REM(TM) Fault Indicator.
REM(TM) (Return Electrode Monitoring) (Standard)
300 Watts Output – Cut, 120 Watts – Coagulation, 70 watts – Biopolar
Monopolar Foot Switch Included
Bipolar Foot Switch (Optional)
Rolling Carts (Optional)
Specifications
Dimensions: 8 x 11 x 17 inches (15 x 28 x 3 cm)
Weight: 15 lbs (6.4 kg)
Output Characteristics:
Cut – 2400 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 200 Nominal Power at rated load watts, 1.8 @ 100 W Crest Factor at rated load typical
Blend 1 – 2800 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 175 Nominal Power at rated load watts, 2.7 @ 100 W Crest Factor at rated load typical
Blend 2 – 3200 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 150 Nominal Power at rated load watts, 3.3 @ 100 W Crest Factor at rated load typical
Blend 3 – 3600 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 125 Nominal Power at rated load watts, 4.6 @ 100 W Crest Factor at rated load typical
Coag – 5000 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 75 Nominal Power at rated load watts, 8.0 @ 50 W Crest Factor at rated load typical
Bipolar – 800 Max (open circuit) P-P Voltage, 100 Rated Load ohms, 50 Nominal Power at rated load watts, 1.8 @ 40 W Crest Factor at rated load typical
Output Waveform
Cut – 500 kHz Sinusoid
Blend 1 – 500 kHz bursts, 50% Duty Cycle, 31 kHz Recurrence
Blend 2 – 500 kHz bursts, 37.5% Duty Cycle, 31 kHz Recurrence
Blend 3 – 500 kHz bursts, 25% Duty Cycle, 31 kHz Recurrence
Coag – 500 kHz bursts, 31 kHz Recurrence
Bipolar – 500 kHz bursts, Unmodulated Recurrence
Low Frequency Leakage:
(50/60 Hz)
Normal polarity: intact chassis ground – <10 mA, ground open - <100 mA.
Reverse polarity: ground open - <100 mA. Sink current, 140V applied, all inputs - <150 mA. Chassis source current, ground open - <110 mA
High Frequency Risk Parameters
Bipolar and Monopolar RF leakage current: <150 milliamperes
REM™ Contact Quality Monitor:
Measurement frequency: 140 kHz ± 10 kHz.
Measurement current: 1.5 mA ± 0.5 mA
Input Power Source:
Nominal input line voltage is 120 VAC with a full range of 85-135 VAC. Frequency line varies from 45 to 65 Hz
Idle - 0.5 A (Max. Current), 60 W (Max. Power)
Cut - 5.0 A (Max. Current), 600 W (Max. Power)
Coag - 30 A (Max. Current), 300 W (Max. Power)
Bipolar - 1.5 A (Max. Current), 180 W (Max. Power)